Self-normalized front photopyroelectric technique for thermal effusivity measurements in liquids

Rumen Ivanov, Ernesto Marin, P. D. Alaniz, J. J. Villa, J. A. Bermejo Arenas

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

5 Citas (Scopus)

Resumen

In this paper a measurement methodology is proposed to avoid some limitations of conventional front photopyroelectric technique for thermal effusivity measurements, related to the necessity of calibration using a reference sample. The same pyroelectric sensor heated at high frequencies acts as a reference sample. The proposed method is tested with measurements of thermal effusivity in liquid samples.

Idioma originalInglés
Páginas (desde-hasta)8-12
Número de páginas5
PublicaciónThermochimica Acta
Volumen599
DOI
EstadoPublicada - 10 ene. 2015
Publicado de forma externa