Quality imaging - Comparison of CR mammography with dcreen-film mammography

E. Gaona, J. Azorín Nieto, J. A. Irán Díaz Góngora, M. Arreola, G. Casian Castellanos, G. M. Perdigón Castañeda, J. G. Franco Enríquez

Producción científica: Contribución a una revistaArtículo de la conferenciarevisión exhaustiva

1 Cita (Scopus)

Resumen

The aim of this work is a quality imaging comparison of CR mammography images printed to film by a laser printer with screen-film mammography. A Giotto and Elscintec dedicated mammography units with fully automatic exposure and a nominal large focal spot size of 0.3 mm were used for the image acquisition of phantoms in screen-film mammography. Four CR mammography units from two different manufacturers and three dedicated x-ray mammography units with fully automatic exposure and a nominal large focal spot size of 0.3 mm were used for the image acquisition of phantoms in CR mammography. The tests quality image included an assessment of system resolution, scoring phantom images, Artifacts, mean optical density and density difference (contrast). In this study, screen-film mammography with a quality control program offers a significantly greater level of quality image relative to CR mammography images printed on film.

Idioma originalInglés
Páginas (desde-hasta)227-229
Número de páginas3
PublicaciónAIP Conference Proceedings
Volumen854
DOI
EstadoPublicada - 2006
Publicado de forma externa
Evento9h Mexican Symposium on Medical Physics - Guadalajara, Jalisco, México
Duración: 18 mar. 200623 mar. 2006