Resumen
In this paper a photoacoustic method is proposed for specific heat capacity measurements in solids using a heat transmission configuration, where illumination takes place at the reference-sample interface. The method involves the use of a liquid reference sample so that it becomes free of instrumental function influence due to effects dependent on the detection cell, the frequency dependence of the microphone response and of the involved electronics. The method is validated with measurements performed in test samples of different materials.
Idioma original | Inglés |
---|---|
Páginas (desde-hasta) | 1208-1211 |
Número de páginas | 4 |
Publicación | Measurement: Journal of the International Measurement Confederation |
Volumen | 46 |
N.º | 3 |
DOI | |
Estado | Publicada - abr. 2013 |
Publicado de forma externa | Sí |