Abstract
The study of the directional properties of a modified Bonner type spectrometer, are reported in this paper. It was established in a previous publication [1], that a small radial hole on each polyethylene sphere gives rise to a directional sensitivity when used on this apparatus. Now, the directional properties are further improved by adding a thin cylindrical cadmium shell inserted in the radial holes of each sphere. The angular response for a set of six spheres is reported using, both, a fast (non-moderated) and a moderated neutron source provided by the 9Be(d,n)10B reaction by bombarding a thick beryllium target with 4 MeV deuterons.
Original language | English |
---|---|
Pages (from-to) | 267-270 |
Number of pages | 4 |
Journal | Revista Mexicana de Fisica |
Volume | 48 |
Issue number | 3 |
State | Published - Jun 2002 |
Externally published | Yes |
Keywords
- Angular response
- Bonner spheres
- Neutron spectrometer