Thermoluminescent response of CaSO<inf>4</inf>:Dy+PTFE induced by X-ray beams

L. Hernández, T. Rivera, Y. Jimenez, R. Alvarez, J. Zeferino, A. Vázquez, J. Azorín

    Research output: Contribution to journalArticle

    7 Scopus citations

    Abstract

    The aim of the present work was to evaluate the feasibility of the utilization of CaSO4:Dy pellets for X-ray measurements in a general radiology department. Thermoluminescence (TL) response of CaSO4:Dy+PTFE was compared to the TL response of commercial LiF:Mg,Ti (TLD-100) samples. TL pellets were exposed to X-ray beam from X-ray machine CMR for clinical diagnosis purpose. The calibration curve of CaSO4:Dy+PTFE was obtained and it showed a linear response as a function of absorbed dose in air at the studied dose interval. Despite this fact, this material can be used for X-ray beams measurements if appropriate calibration procedures are performed. © 2011 Elsevier Ltd.
    Original languageAmerican English
    Pages (from-to)1307-1309
    Number of pages1176
    JournalApplied Radiation and Isotopes
    DOIs
    StatePublished - 1 Jul 2012