Monte Carlo simulation and measurements of clinical photon beams using LiF:Mg,Cu,P+PTFE

C. Azorín-Vega, T. Rivera-Montalvo, J. Azorín-Nieto, L. Villaseñor-Navarro, P. Luján-Castilla, H. Vega-Carrillo

Research output: Contribution to journalArticle

3 Scopus citations

Abstract

Thermoluminescent response of LiF:Mg,Cu,P+PTFE under clinical photon irradiation was obtained. Thermoluminescent dosimeters (TLDs) were irradiated for determining entrance surface dose (ESD) in a solid water phantom when using standard clinical adult treatment protocols. A Monte Carlo simulation of photon interaction with matter was performed and absorbed dose determined. ESD calculated by MCNPX code was greater than those determined by direct measurements in phantom. The results obtained open the possibility for using this material as a TLDs in medical accelerators. © 2009 Elsevier Ltd. All rights reserved.
Original languageAmerican English
Pages (from-to)954-956
Number of pages3
JournalApplied Radiation and Isotopes
DOIs
StatePublished - 1 Apr 2010
Externally publishedYes

Fingerprint Dive into the research topics of 'Monte Carlo simulation and measurements of clinical photon beams using LiF:Mg,Cu,P+PTFE'. Together they form a unique fingerprint.

  • Cite this

    Azorín-Vega, C., Rivera-Montalvo, T., Azorín-Nieto, J., Villaseñor-Navarro, L., Luján-Castilla, P., & Vega-Carrillo, H. (2010). Monte Carlo simulation and measurements of clinical photon beams using LiF:Mg,Cu,P+PTFE. Applied Radiation and Isotopes, 954-956. https://doi.org/10.1016/j.apradiso.2009.12.017